To solve the problems as low detection precision and lack of efficient detection methods, a more scientific and accurate method for the spectral responsivity detection of TDI CCD was put forward. Firstly, the video signal processing technique was improved; Secondly, LED array with different narrow-band wavelengths were chosen as the light source to establish a convenient and accurate detection system for the wide-band spectral responsivity; Then, by changing the irradiance that TDI CCD takes over and detecting the output signal, a kind of high-precision detection method was puts forward; Finally, by calculating the corresponding spectral responsivity of TDI CCD for LED of all narrow-band wavelengths, it is drawn the spectral responsivity curve of TDI CCD in the wavelength scope from 400nm to 900nm. Experiment results show this method realizes higher precision over the wide wavelength scope of 400~900nm, and the detection error of spectral responsivity is 0.025.
参考文献
相似文献
引证文献
引用本文
刘妍妍,李国宁,杨晓慧,刘云荣. TDI CCD宽波段光谱响应度的高精度检测[J].半导体光电,2013,34(2):298-300,325. LIU Yanyan, LI Guoning, YANG Xiaohui, LIU Yunrong. High-Precision Detection of Wide-Band Spectral Responsivity of TDI CCD[J].,2013,34(2):298-300,325.