Abstract:In order to obtain the 213.4nm excitation light required by the UV fluorescence sulfur dioxide analyzer, it was studied the parameter optimization of using dispersive ultraviolet filter, and the impact of the entrance and exit slits on the intensity, stability and spectral purity of the output light was systematically analyzed in the equipment and the physical mechanisms were discussed. The results show that the optimal parameters of the equipment can be set as entrance slit of 50μm and exit slit of 40μm, which is of great significance for improving the accuracy, sensitivity, and long-term stability of the analyzer.