A Monte Carlo model was developed and implemented to calculate the characteristics of X-ray induced secondary electron (SE) emission from a CsI photocathode used in X-ray streak cameras. Energy distributions of emitted SEs were investigated with incident X-ray energy ranging from 1~30keV and a CsI thickness ranging from 100 to 1000nm. Simulation results indicate that the ratio of fast electrons (>50eV) increases with the energy of the incident X-ray and the thickness of the CsI photocathode. When the energy of the incident X-ray is 30keV and the thickness of the CsI photocathode is 1000nm, the ratio of the fast electrons achieves 10.8%. However, when the thickness of the CsI photocathode is 100nm and the energy of the incident X-ray is higher than15keV, the ratio of the fast electrons is about 3.4% and does not increase with the energy of the incident X-ray.