Time-correlated Single Photon Counting-based System for Measuring Fluorescence Decay Characteristics of Scintillators and Optimization Research
Author:
Affiliation:

(Key Laboratory of Radiation Physics and Technology of the Ministry of Education,Institute of Nuclear Science and Technology,Sichuan University,Chengdu 610065,CHN)

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    基于飞行时间法(ToF)的液闪探测器是实现聚变点火研究中子 /伽马诊断的关键方法,其性能核心依赖于闪烁体的快衰减特性。为满足闪烁体衰减特性精确测试的需求,文章基于时间相关单光子计数(TCSPC)原理建立了测试系统,并针对单光子信号获取优化了衰减机构,采用“滤片 +多孔光栅 ”结构。测试结果表明,该系统响应时间为 0.84 ns,光强度测量动态范围达104,测量精度为 0.61 ns,单光子信号占比 95%以上。利用该系统对衰减特性差异显著的 BC501液体闪烁体和 CLYC晶体闪烁体进行了验证测试,成功获得了二者的衰减曲线。分析结果显示,BC501闪烁体的衰减常数分别为 3.7 ns和 35.5 ns,与参考值分别相差 0.5 ns和 3.2 ns;CLYC的衰减常数分别为 1.4 ns,58.9 ns和 1 074 ns,与参考值分别相差 0.4 ns,8.9 ns和 74 ns。分析结果证实了测量系统的可靠性,该系统将应用于后续闪烁体样品的测试对比研究。

    Abstract:

    Liquid scintillation detectors based on the time-of-flight method are important for neutron/gamma diagnosis in fusion ignition research,with rapidly decaying scintillators at the core. To meet the testing requirements for scintillator decay characteristics,this study developed a testing system based on the advanced time-correlated single photon counting principle and optimized the attenuation mechanism on the basis of a“filter + porous grating”structure for single photon acquisition. The system response time was found to be 0. 84 ns,with dynamic range of light intensity measurement of 104 and measurement accuracy of 0. 61 ns. Notably,more than 95% of the signal can be attributed to single photons. Validation tests were conducted on BC501 and CLYC scintillators, which differ considerably in terms of decay characteristics,and the resultant decay curves for both scintillators were obtained. The decay constants of the BC501 scintillator were found to be 3. 7 and 35. 5 ns,which differ from the reference values by 0. 5 and 3. 2 ns,respectively. Similarly,the decay constants of the CLYC scintillator were found to be 1. 4,58. 9,and 1 074 ns,which differ from the reference values by 0. 4,8. 9,and 74 ns,respectively. These results confirm the reliability of the developed measurement system and will be applied in subsequent testing and comparison of scintillator samples.

    参考文献
    相似文献
    引证文献
引用本文

张翔宇,韩纪锋,曲国峰,刘星泉,林炜平.测量闪烁体荧光衰减特性的 TCSPC系统及优化研究[J].半导体光电,2025,46(4):638-644. ZHANG Xiangyu, HAN Jifeng, QU Guofeng, LIU Xingquan, LIN Weiping. Time-correlated Single Photon Counting-based System for Measuring Fluorescence Decay Characteristics of Scintillators and Optimization Research[J].,2025,46(4):638-644.