Screening Method for Evaluating Imaging Stability of Cooled Infrared Detectors
Author:
Affiliation:

(1. The 11th Research Institute of China Electronics Technology Group Corporation,Beijing 100015,CHN; 2. The Sixth Representative Office in Beijing of Rocket Force,Beijing 100043,CHN)

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    针对现行国标判据中制冷型红外探测器成像质量稳定性评价方法不完整的问题,提出一种基于邻域响应一致性的稳定筛选方法。通过构建非稳定像元识别模型与量化评价体系,建立以非稳定像元占比为核心指标的红外探测器成像稳定性筛选标准。具体实施过程包括:采用相同校正系数对探测器进行多周期非均匀性校正,通过滑动窗口算法分析各像元与邻域像元的一致性,结合多次开关机测试累积非稳定像元数量实现红外探测器成像稳定性的判别。所提方法以多次测试数据为支撑,剔除测试误差影响,提高了筛选方法准确性。经过对 320×256阵列规格、640×512阵列规格的探测器进行实验验证,该方法对不同阵列规模的红外探测器应用成像稳定性均有很强的识别能力,同时具有一定普适性和有效性。

    Abstract:

    This paper presents a screening method that can identify unstable pixels,based on the consistency of adjacent pixel responses,and can thus address the limitations of the national standard criteria for evaluating the imaging stability of cooled infrared detectors. The imaging stability of infrared detectors is assessed by counting the number of unstable pixels. This method applies the same non-uniformity correction coefficient to multiple test datasets and analyzes the consistency of each pixel and its adjacent pixels using a sliding window approach. The imaging stability of the detector is determined by accumulating the number of unstable pixels over multiple power on/off cycles. Supported by extensive test data,this approach mitigates test errors and improves the screening method accuracy. Experimental verification of detectors with 320 × 256 array and 640 × 512 array specifications reveals that the proposed method exhibits strong recognition ability,showing potential for ensuring the imaging stability of infrared detectors with different array scales. In addition,the universality and effectiveness of the proposed method are verified.

    参考文献
    相似文献
    引证文献
引用本文

于元祯,高璐,陈彦冠,王亮,邢艳蕾,王晓龙,张磊,李娟.红外探测器成像稳定性筛选方法[J].半导体光电,2025,46(6):1058-1063. YU Yuanzhen, GAO Lu, CHEN Yanguan, WANG Liang, XING Yanlei, WANG Xiaolong, ZHANG Lei, LI Juan. Screening Method for Evaluating Imaging Stability of Cooled Infrared Detectors[J].,2025,46(6):1058-1063.

复制
分享
相关视频

文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2025-07-23
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2025-12-26
  • 出版日期:
文章二维码