Abstract:Laser wavelength is a fundamental metrological reference in precision optical systems and directly limits the accuracy of optical systems and measurement techniques. Single-frequency lasers operating at specific wavelengths are widely used in optical communications, spectroscopy and optical testing techniques. Therefore, accurate wavelength determination is critical. This study designs and implements a compact fiber-optic wavemeter based on the principle of optical path scanning interferometry. The system adopts a Michelson interferometer configuration, where optical path scanning is realized by a stepper-motor-driven mechanism, and a real-time measurement platform is established by integrating a low-noise photodetector with a high-speed data acquisition module. To improve measurement accuracy and long-term stability, engineering optimizations are performed on the interferometric signal acquisition chain, mechanical resonance suppression, and the signal-processing workflow. Experimental results demonstrate that the proposed system achieves a wavelength measurement accuracy of ±2 pm within the 1550 nm telecommunication band. Under continuous measurement conditions, the wavelength drift is less than ±1 pm.