    摘要:

    大面阵全帧转移电荷耦合器件(CCD)的阵列规模较大,其输出图像会存在部分坏像元,并且CCD探测器中广泛存在Brighter-fatter效应,都将影响由图像数据计算的时域方差值和灰度均值,导致器件光子转移曲线的测试不准确,严重影响器件参数评价的精准度。为提高大面阵全帧转移CCD探测器光子转移曲线非线性的评价准确性,本文研究了图像处理和校正方法,通过对图像数据分块、合并、剔除坏像元,校正坏像元引起的图像时域方差变小问题,减小Brighter-fatter效应引起的相邻像元相关性,提高器件光子转移曲线的非线性和多通道增益的准确度。处理结果表明:器件光子转移非线性平均提高97%,同一只器件各通道增益经过图像处理均修正在同一增益范围内。该算法提升了大面阵全帧转移CCD器件的测试准确度和测试效率,对超高分辨率科学仪器用探测器研究具有重要的应用价值和指导意义。

    Abstract:

    Large-format full-frame transfer charge-coupled devices (CCD) feature a large array size, and their output images usually contain defective pixels. In addition, the brighter-fatter effect widely exists in CCD detectors. Both factors affect the temporal variance and gray-scale mean calculated from image data, resulting in inaccurate measurement of the device photon transfer curve and seriously reducing the accuracy of device parameter evaluation.To improve the evaluation accuracy of the nonlinearity of the photon transfer curve for large-format full-frame transfer CCD detectors, this paper studies image processing and correction methods. By partitioning, merging image data and eliminating defective pixels, the problem of reduced temporal variance of images caused by defective pixels is corrected, the correlation between adjacent pixels induced by the brighter-fatter effect is reduced, and the accuracy of the nonlinearity of the photon transfer curve and multi-channel gain of the device is improved.The results show that the photon transfer nonlinearity of the device is improved by an average of 97%, and the gains of each channel of the same device are corrected to the same gain range after image processing. The proposed algorithm enhances the measurement accuracy and efficiency of large-format full-frame transfer CCD devices, which has important application value and guiding significance for the research on detectors used in ultra-high-resolution scientific instruments.

    参考文献
    相似文献
    引证文献
引用本文
分享
相关视频

文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2026-03-17
  • 最后修改日期:2026-04-24
  • 录用日期:2026-04-27
  • 在线发布日期:
  • 出版日期:
文章二维码