    摘要:

    EMCCD因其独特的电子倍增机制而成为光子级微弱光信号探测与成像的关键部件,但EMCCD的倍增增益受工作温度影响较大,退化严重时甚至无法实现极微弱光条件下的探测成像。为了探索器件增益随温度和倍增电压的变化规律并建立增益的标定与校正参考方法,本文采用自研240×240像元规模的高帧频EMCCD器件,通过系统测试和分析该器件在不同温度与倍增电压条件下的增益特性,研究了器件的增益-温度特性关系,拟合出增益经验公式。实验表明:基于电离率温度场理论和倍增增益经验模型,利用测试数据拟合得到关键参数B=383.1、C=0.258及β=1.1975×10-8,建立了该器件的增益-温度-电压经验关系式。该研究可为EMCCD相机的增益标定与矫正提供依据,有助于根据应用场景选取最佳增益倍数,以实现最佳信噪比和成像质量,对自适应光学系统的高速高灵敏波前探测具有重要的应用价值和意义。

    Abstract:

    Electron-Multiplying Charge-Coupled Devices (EMCCDs) are critical components for photon-level ultra-weak light signal detection and imaging owing to their unique electron and imaging owing to their unique electron multiplication mechanism. However, the multiplication gain of an EMCCD is highly susceptible to operating temperatures, and severe gain degradation can even render detection and imaging under extremely low-light conditions unfeasible. To investigate the variations in device gain with respect to temperature and multiplication voltage, as well as to establish a reference method for gain calibration and correction, this study utilizes a self-developed 240×240-pixel high-frame-rate EMCCD. Through systematic testing and analysis of the device’s gain characteristics under various temperatures and multiplication voltages, the gain-temperature relationship is investigated, and an empirical gain formula is subsequently fitted. Experimental results demonstrate that based on the temperature-dependent impact ionization rate theory (ionization rate temperature field theory) and an empirical multiplication gain model, the fitting of experimental data yields key parameters: B=383.1, C=0.258 and β=1.1975×10-8. Consequently, an empirical gain-temperature voltage relationship for this device is established. This study provides a basis for the gain calibration and correction of EMCCD cameras, facilitating the selection of the optimal multiplication gain factor tailored to specific application scenarios to achieve the best possible signal-to-noise ratio (SNR) and image quality. This work holds significant application value and implications for high-speed, high-sensitivity wavefront detection in adaptive optics systems.

    参考文献
    相似文献
    引证文献
引用本文
分享
相关视频

文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2026-07-08
  • 最后修改日期:2026-07-08
  • 录用日期:2026-07-15
  • 在线发布日期:
  • 出版日期:
文章二维码